HI 8733 has been designed specifically for use in areas of production and quality control. It is often necessary to test samples with different concentrations ranging from deionized water to brine.
The amperometric system would require constant recalibration and probe changes, thus making the measurement process very time consuming and unreliable. To solve this problem, HANNA® has developed the HI 8733, which uses a 4-ring potentiometric probe. The rugged probe is made of PVC - ideal for indoor, as well as outdoor measurements.
By using one point calibration, it is now possible to obtain accurate readings across a very wide range, from 0.1 to 199900 µS/cm.
If temperature of the solution fluctuates, HI 8733, with a built-in temperature sensor and Automatic Temperature Compensation, is the perfect instrument.
The HI8733N comes complete as a kit and includes a rugged carrying case with a beaker, a calibration screwdriver and standardizing solution at 12880 µS/cm.
Order Information:
HI 8733 is supplied with HI 76302W probe with internal temperature sensor & 1m (3.3') cable, calibration screwdriver, calibration solutions, protective case, battery and instructions.
Specifications
| Range |
|
0.0 to 199.9 µS/cm;
0 to 1999 µS/cm;
0.00 to 19.99 mS/cm;
0.0 to 199.9 mS/cm |
| Resolution |
|
0.1 µS/cm;
1 µS/cm;
0.01 mS/cm;
0.1 mS/cm |
| Accuracy |
|
±1% F.S. (excluding probe error) |
| Calibration |
|
manual, 1 point |
| Temperature Compensation |
|
automatic, 0 to 50°C (32 to 122°F) with ß adjustable from 0 to 2.5%/°C |
| Probe |
|
HI 76302W, ATC, with 1 m (3.3') cable (included) |
| Battery Type / Life |
|
1 x 9V / approx. 100 hours of continuous use |
| Environment |
|
0 to 50°C (32 to 122°F); RH max 100% |
| Dimensions |
|
164 x 76 x 45 mm (6.4 x 3.0 x 1.8") |
| Weight |
|
250 g (8.8 oz.) |
|